Difference between revisions of "2018-04-10 Cool Down Analysis"

From Nuclear Physics Group Documentation Pages
Jump to navigationJump to search
Line 21: Line 21:
 
| The bypass valve was closed and the heat exchanger was opened slightly.  The roots were off and a secondary small roughing pump was pumping the cryostat space.
 
| The bypass valve was closed and the heat exchanger was opened slightly.  The roots were off and a secondary small roughing pump was pumping the cryostat space.
 
|}
 
|}
[[Image:04-12-2018 magnetT.jpg|500px]]
+
[[Image:04-12-2018 magnetT.jpg|1000px]]
 
 
 
 
  
 
= 04/10/2018 Magnet Precool with LN2 =
 
= 04/10/2018 Magnet Precool with LN2 =

Revision as of 15:01, 23 April 2018

04/12/2018 Fridge Cooldown

The fridge was cool-down starting from room temperature on April 12. The steps that day can be summarized as:

The table's caption
Time Process Comment
10:30 am - 2 pm Cooling down the separator Both needle valves closed and Separator pump flow constant at ~ 5 slm
2 pm - 5 pm Nose cooldown to 1.2 K. The needle valves were opened (both). The roots were on When the temperature in the nose decrease bellow 4 K, the bypass valve was closed and the heat exchanger was opened slightly.
5 pm - 10 pm Thermal equilibrium conditions ~1.9 K. The bypass valve was closed and the heat exchanger was opened slightly. The roots were off and a secondary small roughing pump was pumping the cryostat space.

04-12-2018 magnetT.jpg

04/10/2018 Magnet Precool with LN2

04-10-2018 magnet precool magnetT.png 04-10-2018 magnet precool fridgeT.png 04-10-2018 magnet precool targetT.png 04-10-2018 magnet precool pressure shells.png