2018-04-10 Cool Down Analysis
From Nuclear Physics Group Documentation Pages
Jump to navigationJump to searchThe printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
04/12/2018 Fridge Cooldown
The fridge was cool-down starting from room temperature on April 12. The steps that day can be summarized as:
Time | Process | Comment |
---|---|---|
10:30 am - 2 pm | Cooling down the separator | Both needle valves closed and Separator pump flow constant at ~ 5 slm |
2 pm - 5 pm | Nose cooldown to 1.2 K. | The needle valves were opened (both). The roots were on When the temperature in the nose decrease bellow 4 K, the bypass valve was closed and the heat exchanger was opened slightly. |
5 pm - 10 pm | Thermal equilibrium conditions ~1.9 K. | The bypass valve was closed and the heat exchanger was opened slightly. The roots were off and a secondary small roughing pump was pumping the cryostat space. |
10 pm - 12 am | System off. | The needles valve were close and the separator pump was off. |
Note: Due to connection problems the separator flow wasn't reading all the time. The separator was woking from 10:30 am to 10:30 pm, which means that any 0 reading in the separator flow during that time is a false reading.
Magnet Temperatures and Helium lost estimations
Fridge Temperatures during the complete day
Fridge Temperatures during the cooling process
Fridge Temperatures when the nose was at 1.12 K
Fridge Temperatures when the nose was at 1.9 K
Target Temperatures during the complete day
CCS3 and Vapor Pressure measurements
CCS3 and Pumps Flows
Shells Pressure and CCSF1 in the Fridge